Fault Codes
The following table defines each bit for the fault codes returned when
polling register 7022 and 7023 using function code 03. When a value
of 1 is returned for a bit, it indicates that the fault has occurred since
last power on (7022) or is currently active (7023).
Bit
0
Reset, brownout
1
Reset, reset pin
2
Reset, DoBOR
3
Reset, wakeup from LPM5
4
Reset, security violation
5
Reset, supply voltage supervisor low
6
Reset, supply voltage supervisor high
7
Reset, supply voltage monitor low
8
Reset, supply voltage monitor high
9
Reset, DoPOR
10
Reset, watchdog timer timeout
11
Reset, watchdog timer key violation
12
Reset, flash key violation
13
Reset, PLL unlock
14
Reset, peripheral/configuration area fetch
15
Reset, power management key violation
16
Low battery
17
Pulse output overflow
18
Alarm, rate low
19
Alarm, rate high
20
Alarm, total
21
Flash segment 1 invalid
22
Flash segment 2 invalid
23
Maximum input frequency exceeded
24
EEPROM read error on startup
25
Code execution error
26
Flow rate exceeds 20mA setting
27
Temperature Input Fail
28
Pressure Input Fail
29
Flash Segments Table Invalid
30
Spare 2
31
Spare 1