® ® TABLE OF CONTENTS TRMS DIGITAL MULTIMETER MODEL - KM 60-T TITLE PAGE NO. Safety............II. Electrical Specifications......III. Troubleshooting ........Instrument Layout........Test Certificate......... Warranty...........
® ® I. Overview :The 6000 counts is a pocket-sized 3-5/6 bit automatic FEATURES : digital instrument. Is a stable performance, high precision, high í Backlight Display í Flash Light reliability clear readings, and overload protection. With AAA 1.5V í Pulse Output í...
® ® ELECTRICAL SPECIFICATIONS : DC VOLTAGE Accuracy (reading data of % + least significant digits) accuracy Range Accuracy Resolution storage temperature (23.5)°C, Relative Humidity <75%. 0.001 V ±(0.5%rdg + 3dgts) Performance (Note “ ” Indicates that the instrument has 0.01 V this function ±(0.8%rdg + 10dgts)
Page 5
® ® AC VOLTAGE (ACV) DC CURRENT (DC mA) Range Accuracy Resolution Range Accuracy Resolution 0.001 V 10 mA ±(0.8%rdg + 3dgts) 0.01 V ±(1.2%rdg + 10dgts) 100 mA ±(1.2%rdg + 10dgts) 750 V The Maximum measured pressure drop : 600mV Overload Protection : 60mA / 600mA : 600mA/250V Speed Glass Fuse.
Page 6
® ® NOTE : 1.The instrument series connection to the circuit pending to test before the circuit should be the first power off. 2.If there is no concept of the measured current range in advance, the range switch should be transferred t o the highest range, & then according to the display value to the corresponding file;...
Page 7
® ® 2.Turn the range switch to the corresponding DC/AC mA gear, trigger “ SELECT” to select AC m A and then put the meter into the circuit under test. The measured current value and the current polarity of the red test point will also be displayed at the same time.
Page 8
® ® AC CURRENT (mA) highest range, & then according to the display value to Range Accuracy Resolution the corresponding file; such as the screen display “0L” that has exceeded the range, needed to turn the range 0.1 mA ±(1.2%rdg + 10dgts) switch to the appropriate gear.
Page 9
® ® RESISTANCE (W) 2.Turn the range switch Range Accuracy Resolution to the corresponding 1.000 ±(0.8%rdg + 5dgts) DC/ACA position. The default value is AUTO DC6A/10A. Then RANGE MAX/MIN SELECT HOLD SELECT ±(0.8%rdg + 3dgts) connect the meter to 100 W the circuit under test.
Page 10
® ® 2. Set the range switch to NOTE : “ ” grade, trigger the 0.609 1.When measuring low resistance, the table will bring the internal “SELECT” key, select the resistance, in order to obtain accurate readings, you can firstly diode measurement, and record the short-circuit value of the table pen, by testing value connect the test leads to...
Page 11
® ® Overload Protection : 550V DC or AC peak FREQUENCY 1.Insert the black meter pen Range Accuracy Resolution 4.700 into the “COM”socket while the 4.7 mF 0.01 Hz red meter pen into the “V/Ω/Hz” jack; RANGE MAX/MIN SELECT HOLD 2.
® ® SQUARE-WAVE OUTPUT ( INSTRUMENT LAYOUT : The operation is as followers: (above right figure) 1. LCD Display. 1. Rotate the dial to “ ” position; (default start output is 50Hz); 2. MAX / MIN measurement 2. Insert the two input terminals of the device under test into the key, press &...
Page 13
® ® AUTOMATIC STARTUP & SHUTDOWN NCV MEASUREMENT When the instrument stops using for about 15 minutes, the The operation is as follows: (lower right figure) meter will automatically power off to enter the sleep state; To 1. Rotate the dial to the “NCV” position (LCD displays “EF” in the restart the power, dial to the OFF position, turn the knob to the automatic shutdown function.
® ® WARRANTY MUMBAI Each “KUSAM-MECO” product is warranted to be free from defects in material and workmanship under normal use & service. The warranty TEST CERTIFICATE period is one year (12 months) and begins from the date of despatch of goods.