1. GENERAL This instrument is a high performance, 4½ digital bench multi-meter for measuring DC and AC voltage, DC & AC current, Resistance, Capacitance, Hz, hFE, Diodes and Continuity. The max. Voltage measuring can upto 1000V DC or AC peak value, resolution can be 10 V, and current measuring can upto 20A. It also has the function of polarity auto-change, backlight displaying and directly reading.
Page 4
2. TECHNICAL SPECIFICATION Accuracy = ± (reading % + the lowest effective digit) preheating time : 30min. Temperature for accuracy (23 ± 5)°C, R.H.<75%, one year warranted from the production date DC VOLTAGE (DCV) RANGE ACCURACY RESOLUTION ± (0.05% rdg + 3dgts) ±...
Page 6
DC CURRENT (DCA) RANGE ACCURACY RESOLUTION LOAD VOLTAGE 0.01 mV/uA 101 mV/mA ± (0.35% rdg + 10 dgts) mV/mA mV/mA ± (0.8% rdg + 10 dgts) 275 mV/A mV/A Max. Input current : 20A (within 15s) Overload Protection : 2A / 250V fuse, 20A / 250V fuse AC CURRENT (ACA) TRUE RMS RANGE RANGE...
Page 7
RESISTNCE (Ω) RANGE ACCURACY RESOLUTION SHORT-CIRCUIT CURRENT OPEN-CIRCUIT VOLTAGE 0.01 Ω Ω ± (0.15% rdg + 20 dgts) About 260uA Ω kΩ kΩ Ω About 50uA About 1V ± (0.15% rdg + 5 dgts) kΩ Ω About 5uA MΩ Ω About 0.5uA ±...
Page 8
FREQUENCY Range Accuracy Resolution 20Hz ~ 200kHz 0.01Hz ~ 0.1kHz ±(0.1%rdg + 3dgts) 200kHz ~ 2MHz 0.1kHz ~ 0.001MHz 2MHz ~ 20MHz 0.001MHz ~ 0.01MHz Input Sensitivity : 1.5V RMS ≤10V RMS, the reading value is zero for the signal below 3Hz Overload Protection : 250V DC or AC peak value (within 15s) hFE MEASURING...
The meter is driven by 220V / 110V AC, when operating, connect the power socket to power plug firstly then 4. OPERATION turn on the power. Each meter is equipped with one pair of testleads as standard attachment, to insert the proper jack in front panel is banana plug, connect the pointed end of the testlead to the circuit under test.
Page 11
4.4 ACA : 1. Connect the black test lead to “COM” terminal and the red one to “mA” terminal (max.2A) or “20A” terminal. 2. Set the knob to ACA range, connect the test leads across to the circuit under measured, the polarity will be displayed with the voltage reading value.
Page 12
NOTE : 1. If the current under measure is unknown beforehand, start from the highest range and work down. 2. If only the MSD displays “ ”, it means overrange, should set to a higher range. 3. The max. Input current is 2A for “mA” and 20A for “20A”, large current will blow the fuse. Be careful when measuring 20A range, large current will make the circuit heat, even make the circuit damaged.
NOTE : 1. Do not input voltage or current signal at Capacitance range. 2. Release the capacitor completely to avoid damaging the meter. 3. It should take a few time to be stable when input at 200mF range. 4. The meter may not return to zero at capacitance range, it’s normal and has no effect on the accuracy. 4.9 TRIODE hFE MEASURING 1.
This Test Certificate warranties that the product has been inspected and tested in accordance with the published specifications. The instrument has been calibrated by using equipment which has already been calibrated to standards traceable to national standards. MODEL NO. KM 8045 SERIAL NO. ___________ DATE: ___________ ISO 9001 REGISTERED ...
“KUSAM-MECO”. “KUSAM-MECO’s” warranty obligation is limited, at option, free of charge repair, or replacement of a defective product which is returned to a “KUSAM-MECO” authorized service center within the warranty period.