1260-100/101 Adapt-a-Switch Carrier Module User Manual
Racal Instruments  1998
3. An interrupt is generated to notify the Option-01T that a
reset event has occurred. This informs the Option-01T that
the relays have opened, so that it remains up-to-date on the
relay status.
4. The relays remain open and do not respond to any
commands from the Option-01T.
5. After the reset condition is removed (the external switch is
returned to its normal position), the relays remain open until
otherwise commanded by the Option-01T.
Other Adapt-a-Switch plug-ins and 1260-Series modules that are
equipped to recognize an emergency reset are configured and
operate the same way as the 1260-120. It is important to note that
a global reset condition opens relays on all plug-ins and
1260-Series modules that are connected to the same Option-01T,
regardless of whether they are equipped with the interface for the
external reset switch.
Using the 1260-100/101 3-3