1260-100/101 Adapt-a-Switch Carrier Module User Manual
MODULE
ADDRESS
SWITCH
Racal Instruments  1998
Option-01T controls. You may set the Option-01T logical address
switches in either of two ways:
1. Set the switch to 255. This is the auto-configure setting, and
indicates to the slot 0 controller that it must assign an address
automatically. When the Resource Manager program is run,
the slot 0 controller will then assign a unique address to the
Option-01T.
2. Set the address manually. Valid settings include any number
from 1
to 254
10
10
that of any other module installed in the same VXIbus
chassis.
Although plug-ins and 1260-Series modules do not require logical
addresses (the Option-01T has one instead), they each have a
module address .
The Option-01T uses module addresses to
distinguish between plug-ins, and between 1260-Series modules.
Each plug-in or module must have a unique module address from
one through twelve, inclusive. Models 1260-100 and 1260-101
assign module addresses as follows:

1260-100 Module Address

The module address DIP switch (see figure 2-2) sets the logical
addresses for all plug-ins in the 1260-100 Carrier. The switch
has two settings:
• 1-6 (closed)
When the switch is set to this position, the
module addresses of the plug-ins are from 1
through 6, with 1 being in the left slot of the top
row.
following pattern:
. Make sure that the address does not match
The plug-ins are addressed in the
Getting Started 2-7