Preparation

Installing the Stimulator

Connection of Cables

[Fig.29]
Take the following precautions when installing the stimulator.
1.
Avoid contact with water or chemicals. Also, avoid direct contact with
mist or humidified air from a humidifier as this may cause malfunction or
reduce the service life of the stimulator.
2.
Keep out of direct sunlight as this may result in an unexpected rise in
temperature, causing malfunction or reducing the service life of the
stimulator.
3.
Make sure that the area is well ventilated. If the temperature inside the
stimulator rises, it may cause malfunction or reduce the service life of the
stimulator.
4.
Do not place blankets or other objects on top of the stimulator.
5.
Avoid placing in an area that is exposed to dust.
6.
The power source must have sufficient capacity to supply the necessary
power requirements of the stimulator. Reduced power resulting from
insufficient capacity may affect the performance of the stimulator or activate
the breaker, causing a power failure.
7.
If any irregularity occurs when the power is turned on, turn the power off
immediately and unplug the power cable from the power outlet. At the
same time, mark the stimulator clearly as "Out of Order" and do not use it
until it has been inspected and repaired.
1.
Ground wire
Connect the supplied ground wire to the ground terminal
and connect the other end to a reliable grounding
conductor.
/Avoid using water pipes as the grounding conductor
as their grounding capability may not be reliable.
/It is prohibited by law to use gas pipes as a grounding
conductor.
2.
Power cable
Insert the supplied power cable with 3-pin plug into the
power socket and insert the other end into an AC wall
outlet.
Always use the supplied 3-pin cable for the
power cable.
If any other power cable is used, it may result
in an electric shock to the operator.
[Note] When the power cable with 3-pin plug is grounded
in a 3-pin outlet with grounding electrode, the
stimulator is automatically grounded. However, to
prevent the effects of a broken ground wire inside
t h e p o w e r c a b l e , d o u b l e g r o u n d i n g i s
recommended using the supplied ground wire.
Preparation   26