Stimulation Waveform Output - ADInstruments ML1001 Owner's Manual

Electronic stimulator
Table of Contents

Advertisement

Stimulation Waveform Output

Name
IPolarity selector key
(
, OFF,
)
JOutput voltage display
(VOLTAGE)
14           Description of Parts
This sets the polarity of the waveform generated by stimulation waveform
setup and the output voltage.
J
K
I
Switches the polarity of the stimulation pulse.
and
negative polarity.
The selection governs both the isolator connector and the output terminals.
Output is 0 V when OFF is selected.
Sets the output voltage of the stimulation waveform. Setting is performed by
the same method as for other settings.
Normally, setting is 0 mV or from 10 mV to 10 V. If ±50 V is selected, setting is
0 mV or from 10 mV to 50 V. (1 mV to 9 mV cannot be set.)
This setting does not affect the mixing signal or modulation signal. DC voltage
in DC mode is also set.
When negative polarity (
as it is as minus voltage. Output precision in relation to the command is within
±5% ±50 mV.
M
L
[Fig.15]
Function
indicates positive polarity
) is selected with the polarity selector key, it is read