Recommended separation distances between portable and mobile RF communications
equipment and the 1001
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufactuer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These gudelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
System Composition for EMC test
This Model 1001 is intended for use in the electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
1101 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the 1001 as recommended below, according to the maximum
output power of the communications.
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2
P
0.12
0.38
1.2
3.8
12
This 1001 is tested to comply with IEC 60601-1-2 :2001 with the following
composition. If any part which is not specified by our company is used, the
EMC specifications might not be satisfied.
1001 Electronic stimulator
Output Code (12pc)
80 MHz to 800 MHz
d = 1.2
P
0.12
0.38
1.2
3.8
12
Units
800 MHz to 2.5 GHz
d = 2.3
P
0.23
0.72
2.3
7.2
23
Cable length
2 m