ADInstruments ML1001 Owner's Manual page 26

Electronic stimulator
Table of Contents

Advertisement

Name
QWaveform modulation key
(MODU)
RDirect current key
(DC)
SAlternate key
(ALTN)
Press this key to modulate the external signal to the waveform created in
stimulation waveform setup described above. (The lamp lights.) Connect the
modulated signal to the MODU terminal on the rear of the stimulator. The
VOLTAGE setting is disabled when this key is pressed.
Any
Set output
MODU input
Output after modulation
The range of the input voltage is within ±10 V. Output is modulated in the same
amplitude as the input signal.
Press this key to create DC stimulation. (The lamp lights.) When this key is
pressed, a DC signal is output, regardless of the setting or action of the stimulation
waveform setup described above. The output voltage is set by VOLTAGE. When
the polarity selector key is set to
set to
, minus- voltage is generated. There is no output when set to OFF.
Stimulation waveforms can be output by alternate polarity.
Both the isolator connector and the output terminals are affected. The polarity
is not switched in relation to the MIXING or MODULATION signal.
Function
MODU button ON
, plus-voltage is generated, and when it is
Description of Parts           17
[Fig.19]

Advertisement

Table of Contents