ADInstruments ML1001 Owner's Manual page 41

Electronic stimulator
Table of Contents

Advertisement

Guidance and manufacture's declaration - electromagnetic emissions
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-3
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacture's declaration - electromagnetic immunity
Immunity test
Electrodatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: Ut is the AC mains voltage prior to application of the test level.
This Model 1001 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the 1001 should assure that it is used in such an
environment.
Compliance
The 1001 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
Group 1
any interference in nearby electronic equipment.
The 1001 is suitable for use in all establishments,
Class B
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Complies
Complies
This Model 1001 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the 1001 should assure that it is used in such an
environment.
IEC 60601 test level
6 kV contact
8 kV air
2 kV for power supply
lines
lines
1 kV differential mode
2 kV common mode
<5% Ut (>95% dip in Ut)
<5% Ut (>95% dip in Ut)
for 0.5 cycle
for 0.5 cycle
40% Ut (60% dip in Ut)
40% Ut (60% dip in Ut)
for 5 cycles
for 5 cycles
70% Ut (30% dip in Ut)
70% Ut (30% dip in Ut)
for 25 cycles
for 25 cycles
<5% Ut (>95% dip in Ut)
<5% Ut (>95% dip in Ut)
for 5s
for 5s
3 A/m
Electromagnetic environment guidance
Compliance level
Floors should be wood, concrete or
6 kV contact
ceramic tile. If floors are covered with
8 kV air
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
2 kV for power supply
typical commercial or hospital
environment.
1 kV differential mode
Mains power quality should be that of a
2 kV common mode
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
If the user of the 1001 requires
continued operation during power mains
interruptions, it is recommended that the
1001 be powered from an
uninterruptible power supply or a
battery.
Power frequency magnetic fields should
3 A/m
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Electromagnetic environment
-Guidance
Specifications   32