3 Assembly

3 Assembly

3.1 Requirements

3.1.1 Installation/setup room

3.1.2 System requirements

3.1.3 Monitor

3.2 Installation

3.2.1 Setting up unit

20
Planmeca ProScanner 2.0
NOTE
Only qualified specialists or employees trained by Planmeca are permitted to
install, connect and start using the unit.
The room chosen for set up must fulfil the following requirements:
Closed, dry, well-ventilated room
It should not be a room made for another purpose (e.g. boiler room or
wet cell)
Max. light intensity 1000 Lux, no direct sunlight at the place of installation
of the unit
There should be no large fields of interference (e.g. strong magnetic
fields) present that can interfere with the correct operation of the unit
Refer to the requirements for environmental conditions in section
"Technical data" on page 10
NOTE
The system requirements for the computer systems can be found at
www.planmeca.com/software/specifications/system-requirements.
The monitor must comply with the requirements for digital X-ray imaging with
a high light intensity and wide contrast range.
Strong ambient light, sunlight falling directly onto the monitor and reflections
can make it harder or even impossible to perform a diagnosis based on the
X-ray images.
WARNING
Risk of damage to sensitive components in the unit as a result of shocks or
vibrations.
Do not expose the unit to any strong vibrations or shocks.
Do not move the unit during operation.
Portable and mobile HF communication appliances can interfere with the
effectiveness of electrical medical devices.
Do not stack the unit next to or together with other appliances
If the unit is operated next to other units or stacked with other units,
monitor the unit carefully in the configuration selected in order to ensure
normal operation