Precautions When
3.2
Assembling
To reassemble the KOBRA® CELL it is important to
tighten the locating screws at opposite corners
before tightening the middle screws. This will
ensure that the KOBRA® CELL is correctly
assembled and will not leak.
Do not over tighten locating screws as this will
damage the threads.

Assembly

3.3
Topside view of all
KOBRA® CELL electrodes,
membranes and spacers.
Smallest fitting hole for location peg on left
side.
Largest fitting hole for location peg on right
side.
Corner‚ cut out' on bottom right.
Fig 2. Assembly of KOBRA
CELL
®
Never use metal nuts or ferrules directly with the
KOBRA® CELL.
Use a guillotine tubing cutter to prevent distortion
of the diameter of the PEEK
assembling the KOBRA® CELL.