Error Light Observed On Power Pack; Leakage From Kobra® Cell; Kobra® Cell Causing High Pressure On The Hplc; Storage Plug Broken Off Or Lodged In Kobra® Cell - R-Biopharm KOBRA CELL Instruction Manual

Table of Contents

Advertisement

Error Light Observed on
7.4
Power Pack
The LED will light up if there is no KOBRA® CELL
attached to the current source.
The LED will also light up when the KOBRA® CELL
is dry as the flow of the current source is inhibited.
Check the red and black terminal leads are
connected to the KOBRA® CELL correctly.
Check the KOBRA® CELL has mobile phase running
through it. It is important to keep liquid in the
KOBRA® CELL at all times in order for it to function.
It is advised to run mobile phase through the
KOBRA® CELL for at least 30 minutes prior to
injection of the standards or samples.
Check the power pack. If a connection or
component of the power pack is faulty the power
pack should be replaced.
Leakage from KOBRA® CELL
7.5
The locating screws on the KOBRA® CELL have not
been tightened
correctly. Please refer to 3.2,
Precautions When Assembling.
Check the locating screws on the KOBRA® CELL
have not been over tightened causing damage to
the threads. Once the threads are damaged a new
white plastic housing is
Check the PEEK
tubing is not blocked or
TM
distorted. Replace if required.
KOBRA
CELL Causing High
®
7.6
Pressure on the HPLC
Check the PEEK
tubing is not blocked or
TM
distorted. Replace if required.
Check the inlet and outlet connections that the
PEEK
tubing is inserted into. If they are blocked
TM
they can be unscrewed and sonicated.
Check the internal components in the
KOBRA® CELL have been inserted in the correct
orientation. Reassemble if required (Fig. 1 and 2).
Storage Plug Broken Off or
7.7
Lodged in KOBRA
Return to R-Biopharm Rhône for repair.