Blood Glucose Monitoring System
GT333
User's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RIGHTEST GT333 and is the answer not in the manual?

Questions and answers

Summary of Contents for BIONIME RIGHTEST GT333

  • Page 1 143x100mm Blood Glucose Monitoring System BIONIME CORPORATION No. 100, Sec. 2, Daqing St., South Dist., BIONIME GmbH Taichung City 40242, Taiwan Tramstrasse 16 Tel: +886 4 2369 2388 GT333 9442 Berneck Fax: +886 4 2261 7586 Switzerland User’s Manual E-mail: info@bionime.ch http: //www.bionime.com...
  • Page 3 RIGHTEST GT333 Blood Glucose Monitoring System is intended for self testing outside the body ( in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
  • Page 4 - Please perform the quality control test regularly to make sure the test results are accurate. laboratory measurement on the same blood sample. - RIGHTEST GT333 Blood Glucose Meter can only be used with RIGHTEST GT333 Blood Glucose - The range of blood sample size please refer to strip insert.
  • Page 5 Limitation Limitation - Blood glucose meter readings may be significantly lower than "true glucose levels" in a - Not for screening or diagnosis of diabetes mellitus. hyperglycemic hyperosmolar state, with or without ketosis. - Not for use on critically ill patients, patients in shock, dehydrated patients or hyper- - Caution is advised when glucose values are below 50 mg/dL or above 250 mg/dL.
  • Page 6: Table Of Contents

    Turning On / Off the Meter........................Bluetooth function..........................Caring for Your Meter Auto Code / Detecting Limit........................ Maintain the Products........................Handling RIGHTEST GT333 Blood Glucose Test strip................ Error Messages and Troubleshooting....................Rightest CARE App..........................Testing Procedure Getting Ready for Testing........................
  • Page 7 RIGHTEST Blood Glucose Monitoring System Contents * Your RIGHTEST GT333 Blood Glucose Monitoring System consists of several items. Please GT333 identify each item, learn its name and how it is used. Below are the items included in your RIGHTEST Blood Glucose Monitoring System: *...
  • Page 8 RIGHTEST Blood Glucose Meter RIGHTEST Blood Glucose Meter Test Strip Port Display Window Battery Cover Main Button ( Meter Battery Meter label Left Button ( Right Button (...
  • Page 9 RIGHTEST Blood Glucose Meter RIGHTEST Blood Glucose Meter Indicates a test result stored in memory Indicates when to apply the blood sample Indicates the average result Appears after you insert test strip into meter Indicates a control solution test result Appears when an error occurs Current date under time mode or testing Unit of test result...
  • Page 10 RIGHTEST Blood Glucose Test strip RIGHTEST Blood Glucose Test strip RIGHTEST GT333 Blood Glucose Meter can only be used with RIGHTEST GT333 Blood Glucose Test strip and RIGHTEST Control Solution GC570. The use of other test strips or control solutions can lead to incorrect results.
  • Page 11 Meter Activation and Battery Change Meter Activation and Battery Change Your RIGHTEST Blood Glucose Meter comes with 2 AAA batteries installed. New batteries will provide power to perform approximately 1000 tests under normal use with one data transfer. Press the main button or insert a strip to activate your meter. 1.
  • Page 12 Set Up Your Meter - Set Up the Date, Time and Volume Set Up Your Meter - Set Up the Date, Time and Volume You can enter the Setting Mode two ways. 1. Year setting With the year format blinking, press the Left or Right button until you see 1.
  • Page 13 Set Up Your Meter - Set Up the Date, Time and Volume Set Up Your Meter - Set Up the Date, Time and Volume 4. Time format 12/24H selection 7. Volume Setting With the volume blinking, press the Left or Right button With the time format blinking, press the Left or Right button to switch between 12H and 24H.
  • Page 14 Turning On / Off the Meter Bluetooth function RIGHTEST Blood Glucose Meter Bluetooth default is with " " symbol when mobile App turn on, it will automatically 1. How to turn on connect with your mobile device after user finished the pairing. Once connected Press the Main button or Insert a test strip.
  • Page 15 Bluetooth function Auto Code / Detecting Limit Auto Code NOTE The RIGHTEST Blood Glucose Meter will automatically 1. The meter will automatically turn off the Bluetooth after all the blood glucose readings detect the Code number on the strip. You do not have to Auto code have synced.
  • Page 16 Handling RIGHTEST Blood Glucose Test strip Handling RIGHTEST Blood Glucose Test strip How to handle RIGHTEST Blood Glucose Test strip. Removing RIGHTEST Blood Glucose Test strip: 1. Hold RIGHTEST Blood Glucose Test strip as shown. Inserting RIGHTEST Blood Glucose Test strip : 1.
  • Page 17 - RIGHTEST GT333 Blood Glucose Meter 2. Pull the two parts in opposite directions to remove - RIGHTEST GT333 Blood Glucose Test strip the adjustable depth cap. (Please check the expiration date on the test strip 3.
  • Page 18 Performing a Blood Glucose Test Performing a Blood Glucose Test 7. Hold the base in one hand and pull on the plunger with 12. After inserting a test strip, the " " icon will flash on the other hand. The device will be cocked. Release the the screen.
  • Page 19 Performing a Blood Glucose Test Performing a Blood Glucose Test 16. Pull off the depth adjustable cap. Without touching the CAUTION used disposable lancet, insert the lancet tip into the - Do not apply your blood drop to the sample port on the strip until you see the " "...
  • Page 20 Setting Markers Alternative Site Testing According to actual life status, you can set meal markers with test result. Alternative site testing: palm or forearm blood sampling Palm You have to perform meal marker setting immediately after the meter displays the reading of your test result.
  • Page 21 Alternative Site Testing Alternative Site Testing CAUTION CAUTION - Test results may vary if blood samples are taken from different sites or under certain - Alternate site testing should only be done during steady-state times (when glucose conditions where glucose levels are changing rapidly such as: following a drink, a meal, is not changing rapidly).
  • Page 22 View Window Appearance Understanding Test Results and Messages Make sure your blood sample covers the Blood glucose test results are shown on the meter as mg/dL or mmol/L, depending on the preset of your meter. whole area of the view window to get an accurate test result.
  • Page 23: About Quality Control Testing

    What is a Quality Control Test ? To ensure proper monitoring function, it is necessary to regularly perform a quality control test. Use one of the RIGHTEST Control Solution GC570 when testing your RIGHTEST GT333 Blood If the test result is below 10 mg/dL , "...
  • Page 24: Rightest Blood Glucose Meter

    - RIGHTEST GT333 Blood Glucose Meter - Wipe the bottle cap with a clean tissue before tightly closing the bottle of control solution. - RIGHTEST GT333 Blood Glucose Test strip - Close the bottle of control solution tightly immediately after using.
  • Page 25: Performing A Quality Control Test

    Performing a Quality Control Test Performing a Quality Control Test 1. Take one RIGHTEST Blood Glucose Test strip from the vial and 9. Tightly replace the cap on the RIGHTEST Control Solution bottle. CLOSE 10. The control solution result will appear. Compare your Quality Control Test result to close the vial cap immediately.
  • Page 26: Understanding Control Test Results

    Understanding Control Test Results Recalling Test Results Your control solution test results should fall within the control solution range printed on the test strip RIGHTEST Blood Glucose Meter is able to automatically store a maximum of 500 test results with vial label.
  • Page 27 Recalling Average Test Results Recalling Average Test Results RIGHTEST Blood Glucose Meter provides several average test results. View the 1-day, 7-days, 5. Quick Searching: 14-days, 30-days, 60-days and 90-days average test results for better blood glucose monitoring. To view test results in sequential order automatically, enter the Memory Mode. Press the Right or Left button for 2 seconds.
  • Page 28: Maintain The Products

    Maintain the Products Maintain the Products Maintenance The cleaning procedure is to remove dust, blood and body fluid from the surface and should be performed whenever the meter or lancing device is visibly dirty. Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying Performing the cleaning procedure once per week is recommended.
  • Page 29 Maintain the Products Maintain the Products To clean the lancing device: NOTE 1. Throughly wipe the entire surface of the lancing device with - Clean and disinfect the outside of the device only. disinfecting wipes listed above to clean any possible dirt, dust, Do not remove battery cover when cleaning and disinfecting.
  • Page 30: Error Messages And Troubleshooting

    Maintain the Products Error Messages and Troubleshooting Temperature Error CAUTION 1. In order to get accurate testing results, please test between 10-40°C(50-104°F). - Users should wash their hands thoroughly with soap and water after handling the meter, lancing device or test strips. 2.
  • Page 31 Error Messages and Troubleshooting Error Messages and Troubleshooting Battery Error Er1- The inserted test strip has been used or damaged. Please use a new test strip. 1. The " " symbol blinks when the battery power is low. Please change the battery as soon as possible. You can still test. Er2- The meter has malfunctioned.
  • Page 32 "Operation Guide" on the App. work, please contact your local distributor. For more information, please visit the official website http://www.bionime.com.tw CAUTION A blood sample should ONLY be applied to the test strip after the test strip has been inserted correctly and the Meter's screen is showing the image of the test strip and a blood drop flashing symbol.
  • Page 33: Specifications

    Specification Specification Measurement Technology Oxidase Electrochemical Sensor Power Supply 2 AAA Batteries Approximately 1000 tests, each test with one data Battery Life Sample Capillary whole blood transfer Meter Dimension 85mm x 57mm x 22.5 mm Measuring Range 10 - 600 mg/dL Meter Weight 81.0 ±...
  • Page 34: Warranty

    Warranty Customer Service Bionime Corporation warrants that this product will be free from defects in materials and workmanship Please review all of the instructions to make sure you are performing the steps correctly. If you have for five years from the date of purchase.
  • Page 35: Expected Blood Glucose Values Without Diabetes

    No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan Fasting <100 Product complies with in vitro Diagnostic Medical Device Directive 98/79/EC. ( CE0197 ) EC-Rep: Bionime GmbH, Tramstrasse 16, 9442 Berneck , Switzerland Two Hours after meals <140 References...