PLEASE NOTE: THIS EQUIPMENT HAS BEEN DESIGNED TO PROVIDE RELIABLE SERVICE;
HOWEVER, BEFORE ATTEMPTING TO INSTALL, OPERATE OR SERVICE THE EQUIPMENT, THIS
INSTRUCTION MANUAL MUST BE READ, UNDERSTOOD AND OBSERVED. FAILURE TO DO SO CAN
RESULT IN IMPROPER OPERATION WITH POSSIBLY HAZARDOUS CONSEQUENCES.
Book # 52090680
Instruction Manual
Absolute Pressure Gauge
Series 1500
Series 1000
Series 300
WIKA Instrument Corporation
1000 Wiegand Boulevard
Lawrenceville, GA 30043
1-888-945-2872
http://www.wika.com
Page 1

Advertisement

loading
Need help?

Need help?

Do you have a question about the 1500 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for WIKA 1500 Series

  • Page 1 WIKA INSTRUMENT CORPORATION Instruction Manual Absolute Pressure Gauge Series 1500 Series 1000 Series 300 WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-945-2872 http://www.wika.com PLEASE NOTE: THIS EQUIPMENT HAS BEEN DESIGNED TO PROVIDE RELIABLE SERVICE; HOWEVER, BEFORE ATTEMPTING TO INSTALL, OPERATE OR SERVICE THE EQUIPMENT, THIS INSTRUCTION MANUAL MUST BE READ, UNDERSTOOD AND OBSERVED.
  • Page 2: Safe Operation And Maintenance

    NOTE Minor part number changes may be incorporated into WIKA products from time to time that are not immediately reflected in the instruction book. If such a change apparently has been made in your equipment and does not appear to be reflected in your instruction book, contact your local WIKA sales office for information.
  • Page 3: Installation

    INTRODUCTION This instruction book provides installation, operation and maintenance instructions for the WIKA Absolute Pressure Gauges Series 1500, Series 1000 and Series 300. WARNING: TO AVOID POSSIBLE SEVERE PERSONAL INJURY OR DAMAGE TO THE EQUIPMENT, THIS EQUIPMENT SHOULD BE INSTALLED, OPERATED AND SERVICED ONLY BY TRAINED, QUALIFIED PERSONNEL WHO ARE THOROUGHLY FAMILIAR WITH THE ENTIRE CONTENTS OF THIS INSTRUCTION BOOK.
  • Page 4: Calibration Check

    (980.665 cm/sec²) and temperature (O⁰C for mercury columns or 20⁰C for water columns, where applicable). While aneroid instruments are not subject to changes in reading due to gravity they are affected by temperature. WIKA High Precision instruments read directly in terms of standard conditions (see preceding paragraph) at a temperature of 23°C unless other-wise indicated.
  • Page 5 Put the gasket in the groove. Install the special bezel and tighten all screws evenly. Evacuate the instrument for one hour. Push in on the screwdriver to engage the adjusting screw. The pointer will rotate in the direction the screwdriver is turned. After the pointer is set t remove the special bezel and replace the instrument bezel and glass.
  • Page 6: Operation

    ABSOLUTE PRESSURE GUAGE OPERATION SERIES 1500 BELOW 100 P.S.I.A. SERIES 1000 BELOW 100 P.S.I.A. SERIES 300 ALL RANGES DWG: 610.050.170.011 PRESSURE RELIEF VALVE: (ON SERIES 1000 AND 1500 ONLY) (BACK OF CASE) PRESSURE SENSITIVE ELEMENT SEALED AT ZEAO ABSOLUTE PRESSURE SYSTEM PRESSURE APPLIED INTO CASE (DO NOT EXCEEO MAX.
  • Page 7 ABSOLUTE PRESSURE GUAGE OPERATION DWG: 610.050.170.021 Book # 52090680 Page 7...
  • Page 8 Book # 52090680 Page 8...
  • Page 9 Book # 52090680 Page 9...
  • Page 10 Book # 52090680 Page 10...
  • Page 11 Book # 52090680 Page 11...