Hazardous Locations Certifications - Tescom ER3000 Manual

Hide thumbs Also See for ER3000:
Table of Contents

Advertisement

HAZARDOUS LOCATIONS CERTIFICATIONS

Factory Mutual (FM) Approval:
Explosion Proof
Identification No. 2Z0A0.AE
Class I, Division I, Groups B, C, and D
Enclosure type 4X
SPECIAL CONDITIONS FOR SAFE USE
– -20°C ≤ T
– Do not remove cover while circuits are live
– Installation to be in accordance with the latest edition of
National Electrical Code
– Unused conduit entry must be closed with suitable blanking
element
– Seal conduit within 18 inches
Canadian Standards Association (CSA)
Explosion proof
Certification No. LR 85614-4
Class I, Groups B, C, and D
Enclosure Type 4X
SPECIAL CONDITIONS FOR SAFE USE
– -20°C ≤ T
– Do not remove cover while circuits are live
– Installation to be in accordance with the latest edition of
Canadian Electrical Code
– Rated 24 VDC, 350 mAmps
– Maximum working pressure: 110 PSIG
KEMA/Cenelec
Flameproof Certification
Certificate No. KEMA 03ATEX2365
ATEX Marking: EEx d IIB + H
Category: II 2 G
Enclosure type IP65
SPECIAL CONDITIONS FOR SAFE USE
– -20°C ≤ T
– Do not remove cover while circuits are live
– Installation to be in accordance with the applicable local
requirements
– The cable and conduit entry devices shall be of certified
flameproof types, suitable for the conditions of use and
correctly installed
– Unused apertures shall be closed with certified flameproof
blanking elements
4
≤ 60°C
amb
≤ 60°C
amb
≤ 60°C
amb
T4