Available languages

7. Starting the melting point analysis

7.1 In principle
Selected temperature
> oven temperature
After that, the device will return to the starting display
and the procedure will be cancelled. The fan will be
started according to the logic.
The latest selection of temperature can only be erased
when turning off the device.
The starting of a melting point analysis will only be
effected by inserting a filled capillary tube in the oven, if
a starting temperature has previously been selected and
reached.
In the case of an unfilled capillary tube, an error warn-
ing will be displayed. It will be automatically deleted
after the empty capillary tube has been removed.
empty capillary tube
After inserting, measurement will automatically start.
When the capillary tube is inserted, the starting menu is
automatically shown.
If there is already a capillary tube in the oven before the
starting temperature has been selected, a message on
the display will ask you to remove the capillary tube.
Please remove
Only then (automatically after removing the capillary
tube) the selecting menu will be shown.
The last selected starting-temperature remains in
memory until the device is turned off.
7. Starting the melting point analysis
capillary tube
27