Structure Process / Operator Programme - kruss M5000 Instruction Manual

Melting point meter
Hide thumbs Also See for M5000:
Table of Contents

Advertisement

Available languages

Available languages

6. Structure process / operator programme

6. Structure process / operator programme
6.1 Starting display
6.2 Selection of the
target temperature
26
actual temperature
The display always appears when switching on the
device.
If the temperature is above 35 °C, the fan will start
automatically.
If the temperature is below 30 °C, the fan will switch off.
If one of the arrow-keys is pressed, you continue with
6.2 Selection of the target temperature.
select temperature
Adjustment takes place in steps of 1 °C.
The menu can only be left by turning the device off and
on or by time-out.
After 5 min (without pressing any key) time-out
appears and the device will be restarted
(see 6.1 Starting display).
When choosing a temperature, the fan will always be
switched off.
If a starting temperature higher than the current oven
temperature is selected, inserting the capillary tube will
start the normal process of melting point analysis.
If the starting temperature is lower than the current oven
temperature, an error message will be displayed for 3 s
when the capillary tube is inserted.
25.8 °C
120.0 °C

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents