8. Reaching the melting point

8. Reaching the melting point
melting point
121.9 °C auto
A signal tone will sound three times when the melting
point has been reached.
The determined melting point will be shown on the
display. The device will automatically set the melting
oven back to the selected starting temperature. During
this time the term "auto" will flash up in the lower right
corner of the display.
When the starting temperature has been reached once
more, the signal tone will sound and the term "auto"
will remain in the lower right corner of the display.
If a new capillary tube is inserted during the time the
term "auto" is flashing up, no error message will be
displayed. As soon as the starting temperature has been
reached, the next melting point analysis process will
immediately be started.
If a new melting sample is inserted within 5 min, the
measuring process will be repeated.
If one of the two arrow keys is pressed within 5 min, the
selection of the target temperature will be activated (see
chapter 6.2) No fan will be turned on!
If the printing key is pressed within 5 min, you can print
on the optionally attached printer. This process can be
repeated as often as you want.
If no keys are pressed for about 5 min, a time-out
will occur and the device will automatically jump to
the starting display. The last melting point remains in
memory and can be displayed at a later time by press-
ing the printing key.
29