Adding and dispensing process chemicals
Refill indicator
Dispensing
rinse aid
46
Close the container.
Wipe up any spilled rinse aid. This prevents over-foaming occurring
during the next program.
When the fill level is low in the (DOS 2) supply container for rinse aid
you are reminded to refill it.
Confirm the message shown with OK and
refill the rinse aid as described.
The dispensing concentration is set by Miele Service.
- If there are spots of water left on load items after reprocessing, then
the dispensing concentration is set too low.
- If clouding or smearing appears on load items, the dispensing
concentration is set too high.
In either case, contact Miele Service and have the dispensing
concentration reset.
Refill rinse aid