6.3 DC mA Measurement

Range
20mA
-10 ℃~ 18 ℃, +28 ℃~ 55 ℃ temperature coefficient,±0.005%FS/ ℃ .
Input resistance: < 100Ω.

6.4 DC mA Source

Range
20mA
20mA
(Transducer
simulation)
-10 ℃~ 18 ℃, +28 ℃~ 55 ℃ temperature coefficient,±0.005%FS/ ℃ .
Maximum load voltage: 20V, equivalent to voltage of 20mA on 1000Ω load resistance.

6.5 Frequency Measurement

Range
100Hz
1000Hz
Maximum
measurement
range
0~24mA
Maximum
Resolution
output range
0~24mA
0.001mA
0~24mA
0.001mA
Maximum
measurement
range
1~99.999Hz
100~999.99Hz
Accuracy (% of reading + Counts)
Resolution
ETX-1815
0.001mA
Accuracy (% of reading + Counts)
ETX-1815
0.05+2
0.05+2
Resolution
0.001Hz
0.01Hz
ETX-2015
0.05+2
0.02+2
ETX-2015
0.02+2
0.02+2
Accuracy (% of reading +
Counts)
ETX-1815
ETX-2015
0.02+1
0.01+1