Shut Down Mode - East Tester ET-2715 User Manual

Current and voltage calibrator
Table of Contents

Advertisement

7.
Press
until the buzzer works to enter 1V as the 0% value.
8.
Press
to select 5 V for the output value, and then press and hold
until the buzzer work to enter 5V as the 100% value.
9.
Press
The screen will display as shown in Figure 2.1-5.

2.2 Shut Down Mode

The calibrator comes with the shutdown mode enabled for a time duration set to 30
minutes (displayed for about 2 seconds when the calibrator isinitially turned on). When
the shutdown mode is enabled, the calibrator will automatically shutdown after the
time duration has elapsed from the time the last key was pressed. To disable the
shutdown mode, press
and
simultaneously. To adjust the time duration, press
simultaneously, the screen will display as shown in Figure 2.2-1,then press
to adjust the time between 1 and 30 minutes and then press
new time duration (Without pressing any key for 5 seconds, the calibrator will quit
from the adjustment automatically).
to select 1 V for the output value, and then press and hold
or
to step between 0 and 100% in 25% step increments.
Figure 2.1-5
and
simultaneously. To enable the mode, press
and
and/or
to store the
7