Safety and Regulatory Information

The product documentation helps you use the R&S RT-ZP03S safely and effi-
ciently. Follow the instructions provided here and throughout the manual.
Use the product only for its intended use and within its performance limits. If the
probe assembly is used in a manner not specified by the manufacturer, the pro-
tection provided by it may be impaired.
Operating site
Only use the product indoors, and keep it dry. The product has no casing and is
sensitive to moisture and humidity.
Do not use the probe in explosive atmospheres.
The product is suitable for pollution degree 2 environments where nonconductive
contamination can occur. Occasionally a temporary conductivity that is caused by
condensation must be expected. Temporary condensation occurs only when the
product is out of service.
Using the probe
Take the following measures for your safety:
● Before connecting the probe to a circuit under test, make sure the probe is
connected to a grounded measurement instrument.
● Do not connect a probe to any voltage that exceeds the maximum permissible
input voltage specified in the data sheet.
● Do not use the probe for measurements on mains circuit.
● Do not use the probes in circuits with measurement category II, III or IV.
● With increasing frequency the voltage on the probe tip should not be higher
than illustrated in the derating curve.
● Always connect the probe lead to earth ground.
Declaration of Conformity
The manufacturer declares the conformity of this product with the actual required
safety standards in accordance with the Low Voltage Directive IEC 61010-031
Safety requirements for electrical equipment for measurement, control and labo-
ratory use.