Other Operator Fault Analysis; Resetting Faults - GE Proteus XR/a Operator's Manual

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 11
Audio Message
12-2-3
12-2-4 Generator Error Codes
Table 12-2
GENERATOR ERROR CODE

12- 3 Other Operator Fault Analysis

TABLE 12-3
OPERATOR FAULT ANALYSIS
Fault
System circuit breaker on
and LCD display is dark.
No exposure.
Image quality.
Fault Reset (AEC).

12- 4 Resetting Faults

In order to assure safe data saving, an audio message of -Bi- Bi- Bi- sounds
when retrieve data from floppy disk.
During X-ray emission an audio message sounds.
Error Code
Err 30
Err 40
Err 50
Err:60
Err 70
Err 80
Err 90
Err100
Err 110
Err 120
Recommended Action
Verify:
– Power line is plugged into the receptacle or connected to the main circuit breaker?
– Power is available at the receptacle or the main circuit breaker?
– The remote ON/OFF switch on the System Console is in the ON position?
Are any error messages displayed?
– Motion artifacts?
Relax the patient. Make the patient as comfortable as possible. Check once again for
proper positioning.
– Insufficient contrast? Decrease kVp.
Check the dark room, film expiration date, chemicals and developer temperature.
Verify:
– Proper chamber selected?
– Chamber connected?
– Film speed correct?
If the film is too dark, decrease density.
If the film is too light, increase density.
When AEC mode is enabled, exposure time/termination is controlled via
ionization chamber sensing circuits. In addition, a back-up time is programmed
into the system to help prevent excessive patient dose in the event of an AEC
system malfunction. If the exposure time equals the back-up time, exposure will
be terminated and a fault message will appear across the bottom of the screen.
To clear the fault, press the function key by the flashing FAULT RESET field. If
the fault persists, the system should be used in manual mode until the fault can
be isolated/corrected.
Meanings
Tube Spits error
Rotation error
Heater ( filament) error
Exposure error
Power supply error
Hardware error
Software error
System Comminication error
Tube/generator overheat error
Application error
12-4
PROTEUS XR/a
Operator Manual
DIRECTION 2259724-100