Physical Requirements Of Room; Environmental Requirements/Limitations - GE Proteus XR/a Operator's Manual

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 11
CHAPTER 13 PHYSICAL REQUIREMENTS OF ROOM

13-1 Environmental Requirements/Limitations

13-1-1 Room Climate
Relative humidity and temperature
See Table 13-1. To obtain relative humidity and temperature requirements for components
TABLE
13-1
CLIMATE REQUIREMENT (BY COMPONENT) - RELATIVE HUMIDITY AND TEMPERATURE
PRODUCT OR
COMPONENT
Elevating table
OTS
Generator
System Console
Tomolink Option
X-ray Tube (Rad 14)
X-ray Tube (MX 100)
SG120 Wallstand
RELATIVE HUMIDITY
IN-USE
STORAGE
MIN.
MAX.
MIN.
10%
75%
10%
75%
10%
75%
10%
75%
10%
75%
10%
75%
10%
80%
20%
85%
TEMPERATURE
IN-USE
MAX.
MIN.
5%
95%
+10°C
5%
95%
+10°C
5%
95%
+10°C
5%
95%
+10°C
5%
95%
+10°C
5%
95%
+10°C
5%
95%
0°C
10%
95%
+10°C
13-1
PROTEUS XR/a
Operator Manual
DIRECTION 2259724-100
STORAGE
MAX.
MIN.
MAX.
40°C
-20°C
40°C
-20°C
40°C
-20°C
40°C
-20°C
40°C
-20°C
40°C
-20°C
40°C
-40°C
40°C
-20°C
+70°C
+70°C
+70°C
+70°C
+70°C
+70°C
+80°C
+70°C

Hide quick links:

Advertisement

Table of Contents
loading