Technical Data

6
GA09313_0202 – (2008-02) –
2
Technical Data
Measurement range (Air, N
Accuracy (N
)
2
Repeatability
Gas type dependence
Output signal (measuring signal)
Voltage range
Measurement range
Voltage vs. pressure
Error signal
Output impedance
Minimum loaded impedance
Response time (pressure dependent)
-6
p > 10
mbar
-8
p = 10
mbar
Identification transmitter
Status
-2
p > 10
mbar
Pirani-only mode
-2
p < 10
mbar
Cold cathode not ignited
Pirani-only mode
p < 10-2 mbar
Cold cathode ignited
Combined Pirani / cold cathode
mode
LED
Supply
The transmitter may only be connected to power supplies, instruments
or control devices that conform to the requirements of a grounded
extra-low voltage (SELV-E according to EN 61010). The connection to
the transmitter has to be fused
Voltage at the transmitter
Power consumption
1)
Fuse
The minimum voltage of the power supply unit must be increased proportionally to
the length of the sensor cable.
Voltage at the supply unit with maxi-
mum cable length
1)
Oerlikon Leybold Vacuum controllers fulfill these requirements.
¤
Oerlikon Leybold Vacuum
-9
)
5×10
... 1000 mbar
2
≈±30%
in the range 1×10
≈±5%
in the range1×10
→  28
0 ... +10.5 V
1.82 ... 8.6 V
logarithmic , 0.6 V / decade (→  26)
<0.5 V no supply
>9.5 V Pirani sensor defective
2×10 Ω
10 kΩ, short-circuit proof
<10 ms
≈1000 ms
85 kΩ, referenced to supply common
pin 6
Low = 0 V
Low = 0 V
High = 15 ... 30 VDC
High voltage on (LED on)
DANGER
1)
.
15 ... 30 VDC (ripple ≤ 1 V
≤2 W
≤1 AT
16 ... 30 VDC (ripple ≤ 1 V
-8
... 100 mbar
-8
... 100 mbar
(filament rupture)
)
pp
)
pp