Decontamination; Decommissioning - Binder redLINE RE 53 Operating Manual

Table of Contents

Advertisement

8.2

Decontamination

Disconnect redLINE chambers from the power supply prior to decontamination. Pull the power plug.
You can use the following disinfectants:
Inner chamber
Standard commercial surface disinfectants free from acid or halogenides.
Alcoholic solutions.
In case of contamination of the interior by biologically or chemically hazardous material, there are three
possible procedures depending on the type of contamination and charging material.
• The ovens RE and RF can be hot air sterilized at 190 °C / 374°F for at least 30 minutes. All
inflammable goods must be removed from the interior before.
• Spray the inner chamber with an appropriate disinfectant.
Before start-up, redLINE chambers must be absolute dry and ventilated, because explosive gases
may form during the decontamination process.
With every decontamination method, always use adequate personal safety controls.
Thoroughly remove all residues of disinfecting agents from the interior with sterile (RI), deionized water.
If necessary, ventilate the interior thoroughly.
Residual disinfecting agents.
Danger of samples being disturbed.
∅ Any residual disinfecting agents in the chamber may pollute the samples.
9.

Decommissioning

Turn off the redLINE chamber at the main power switch (7). Disconnect the chamber from the power
supply.
When turning off the main power switch (7), the stored parameters remain saved.
• Clean the redLINE chamber and its trays and dry them completely.
• Temporal decommissioning: See indications for appropriate storage, chap. 3.3.
• Final decommissioning: Dispose of the unit as described in chap. 12.2 to 12.3.