5.3

Turning on the unit

Insert the plug into a socket (chap. 4.1) and turn on the chamber at its main power switch (7).
RF: The fan is permanently active, as long as the chamber is turned on at the main power switch,
independent on the current controller function.
After turning on the chamber, the controller runs a 3 second self-test sequence displaying the software
version. Then it shows the main display (actual value display).
During the equilibration phase of an hour after turning on the chamber, undefined temperature conditions
occur within the unit. During this phase, do not place any sample materials in the chamber.
Danger of samples being destroyed.
Load the chamber only after equilibration of temperature and CO
6.

Operating the controller

Controller setting is identical on all redLINE chambers. The temperature controllers only differ in their
temperature range and the temperature display and setting accuracy (RE and RF with degree accuracy,
RI with accuracy of a tenth of a degree).
After turning on the chamber at the main power switch (7), the controller is in the operating mode
"Standby".
Main menu (Standby)
REDLINE RE53
The display shows the actual temperature value and "OFF".
The scrolling message (e.g. "REDLINE RE 53") indicates the chamber type.
6.1

Operating modes

There are the following 3 operating modes:
Standby (Off): The heating is off. The unit approximates ambient temperature.
Automatic Mode (Auto): Normal operation. The unit controls to an entered set-point.
Manual Mode (mAn): Manual control of the heating power in %
Redline Operating Manual RE / RF / RI 12/2010
CAUTION
.
2
page 27/46