Connection To A Suction Plant (Optional) - Binder redLINE RE 53 Operating Manual

Table of Contents

Advertisement

Danger by electrical short-circuit.
Danger of fire.
Only use the chamber connected to a power source fused to at least 10 Amp.
Electrical hazard.
Danger of death.
∅ Do not connect any chamber with a dented or damaged rear panel to the power supply.
If the chamber has a damaged power cable or rear panel, withdraw it from use
immediately, remove the power plug and contact your dealer to have it repaired.
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in
case of danger.
4.2

Connection to a suction plant (optional)

When directly connecting a suction plant the spatial temperature exactitude, the heating-up and the
recovering times and the maximum temperature will be negatively influenced. So no suction plant should
be directly connected to the outgoing air pipe.
Active suction from the oven must only be performed together with extraneous air. Perforate
the connecting piece to the suction device or place an exhaust funnel at some distance to the
outgoing air pipe.
The exhaust duct and its cover will become hot during operation.
Danger of burning.
∅ Do NOT touch the exhaust duct and its cover during operation.
Redline Operating Manual RE / RF / RI 12/2010
WARNING
DANGER
CAUTION
page 23/46