Guidelines For Safe Lifting And Transportation; Storage; Location Of Installation And Ambient Conditions - Binder redLINE RE 53 Operating Manual

Table of Contents

Advertisement

3.2

Guidelines for safe lifting and transportation

Observe the advice for temporal decommissioning (chap. 9).
Sliding or tilting of the unit.
Damage to the unit.
Risk of injury by lifting heavy loads.
Transport the redLINE chamber in its original packaging only.
Secure the redLINE chamber with transport straps for transport.
∅ Do NOT lift or transport a redLINE chamber using the door handle or the door.
Lift a redLINE chamber at its four lower corners with the aid of 2 people and place it on
a transport pallet with wheels. Push the pallet to the desired site and then lift the unit
from the pallet at its four lower corners.
• Permissible ambient temperature range during transport: -10 °C / 14°F to +60 °C / 140°F.
You can order transport packing and pallets for transportation purposes via www.binder-redline.com.
3.3

Storage

Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 9).
• Permissible ambient temperature range during storage: -10 °C / 14°F to +60 °C / 140°F.
• Permissible ambient humidity: max. 70 % r.H., non-condensing
If following storage in a cold location the unit is transferred to the installation site for start-up,
condensation may form. Wait at least one hour until the oven has attained ambient temperature and is
completely dry.
3.4

Location of installation and ambient conditions

Set up the redLINE chamber on a flat, even, and non-flammable surface, free from vibration and in a well-
ventilated, dry location and align them using a spirit level. The site of installation must be capable of
supporting the unit's weight (see technical data, chap. 14.4). The chambers are designed for setting up
inside a building (indoor use). Set up the chamber in such a way that the main power switch (7) is easily
accessible and can be turned off immediately in case of danger.
Danger of overheating.
Damage to the unit.
∅ Do NOT set up redLINE chambers in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
• Permissible ambient temperature range during operation: +18 °C / 64.4°F up to +40 °C / 104°F. At
elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient
temperature of +25 °C / 77°F to which the specified technical data relate. For other
ambient conditions, deviations from the indicated data are possible.
Redline Operating Manual RE / RF / RI 12/2010
CAUTION
CAUTION
page 20/46