Select "Tightness Test"
2.9.7
Smart Probe: t510i
2.9.8
Cooling/ Heating Power
Needed Smart Probes: 2x t605i
1. Press
and select
Edit view
2. Press
and select
configure measurement
Select the maximum allowed pressure drop, stabilisation time,
test time and vibration alarm
3. Select
Pressure setting
4. Select
Zero pressure sensor
5. Above the graph the following data is displayed:
Pressure at the start of the measurement
Current pressure
Delta between actual measured pressure and reference pressure
Maximum allowed pressure drop
6. Press
to start the stabilisation time. As soon as stabilisation
time is over, test time starts. Vibration alarm informs about end of
measurement
 Press
to abort the test
1. Press
and select
Edit view
Press
to select humidity probe t605i for Return air and
Supply air and high-pressure probe
 App automatically calculates enthalpy
2. Press
and select
Configure measurement
Enter volume flow to calculate BTU/h
Select return air or supply air
see "2.5 Settings view"
 see "AC + Refrigeration"