1 Overview Thank you for purchasing Apera Instruments DO8500 Portable Optical Dissolved Oxygen Meter. The DO8500 measures dissolved oxygen in water using luminescence technology through an optical sensor and displays data with intelligent instrumentation. Compared to conventional electrochemical dissolved oxygen meter, the DO8500 is more accurate, stable and easier to use.
Instructions 3.1 LCD Screen Fig.-1 Serial number of storage and Measurement mode ① ⑨ prompts of special display mode M+ — Measurement to be stored Reading/Measured Value ② ⑩ RM — Reading to be recalled Temperature value and Timing storage ③...
3.2 Key Operation Fig.-2 Short press: key press time < 2 s; Long press: key press time> 2 s. Power on: Press to turn on. Shutdown: long press 2 seconds off. First time use: see parameter setting P4.7 & P4.8 to set correct date and time (as Clause 8.3) Special notes: Before getting readings or performing other operations, wait about 30 seconds after meter is powered on.
3.3 Batteries The instrument uses three AA alkaline batteries. Battery life > 200 hours (without backlight). When the display shows symbol as shown in Figure-3, replace the battery. 3.4 Instrument Socket Fig.-3 The instrument sockets are protected by grey rubber sealing cap, as shown in Figure-4.
3.6 Data storage, recall, and clear. 3.6.1 Manual storage When the measurement is stable, press , the meter will store the measurement at the moment, M+ icon and storage serial number Fig.-9 will be displayed on the botton right of sreen, as shown in Figure-9, First group of data was stored.
to turn on auto power off function, the instrument will shut down automatically if no operation within 20 minutes, select Off to disable this function, meaning the meter won’t power off until the user manually turn it off. 4 Optical Dissolved Oxygen Probe 4.1 Probe Structure The DO803 optical dissolved oxygen probe has a cable length of 3m and a built-in temperature sensor for automatic temperature compensation.
(C) Before the first use, unscrew the calibration sleeve to check if the sponge is damp. If the sponge is dry or if the electrode is exposed to dry air for more than 8 hours, the surface coating of the sensor cap may be dried out. So the electrode should be soaked in clean water at room temperature for 24 hours.
switch between mg / L and %, or ppm → %. Users can choose mg / L or ppm in parameter setting P3.1, but only percentage (DO%) is displayed in calibration mode. Resolution Selection The resolution unit can be selected in parameter setting P3.2: 0.01 or 0.1mg/L (ppm), After setting, the meter will display resolution of 0.1 or 1 in according to %.
the probe. Tighten the locking cap. Be careful not to have water droplets on the head of the sensor cap. Wait for 5 to 10 minutes after turning on the instrument in order to saturate the air in the calibration sleeve with water vapor. In addition, wait for the temperature to completely stablize.
Note: the brief stirring of the probe in solution is only to eliminate bubbles. Unlike conventional galvanic/polarographic electrodes, the measuring via optical dissolved oxygen probes does not require constant stirring of the solution or flowing fluid. 7.2 Users can read the measurements when appears and stays on.
Page 15
P4.7 Adjust Date P4.8 Adjust Time DO Parameter setting (press to switch) P3.1—Dissolved Oxygen Unit (mg/L—ppm) 1. In P3.0 mode, press to enter P3.1 mode. 2. Press , mg / L flashes, press to select mg/L → ppm, press to confirm. 3.
Page 16
P3.6—Back to Factory Default (No—Yes) Press , No flashes, Press to select No →Yes, Press to confirm. Meter returns to measurement mode. No—cancel restore. Yes—restore Basic Parameter setting (press to switch) P4.1—Adjust timing storage time 1. In P4.0 mode, Press to enter P4.1 mode.
P4.5—Auto Backlight (On—Off) 1. Press , On flashes, Press to select On→Off, Press to confirm. On – turn on auto backlight off function, Off – turn off auto backlight off function. 2. Press to enter P4.6 mode, or press to return to measurement.
Page 18
stored in the meter as shown in Figure – 15. ② — Mode and Serial Number ③ — Parameter setting information ④ — Port type and press key The port number shows where the software is connected to the computer, each computer recognizes a different port number, as shown in Figure –...
Press “Export” key to export the stored value to Microsoft Excel file for further analysis or printing. 10 Complete Kit 10.1 What’s in the box Content Quantity DO8500 Portable Optical Dissolved Oxygen Meter DO803 Optical Dissolved Oxygen Probe Probe Calibration/storge sleeve Conductivity Electrode 2301-3M PC-Link Communication Software Flashdrive USB Communication Cable...
Calibration/Storage sleeve 11 Warranty 11.1 The warranty period of the DO8500 instrument (only the instrument) is 3 years from the date of purchase. That of DO803 optical DO probe (excluding sensor cap) is 2 years from the date of purchase. That of DO8032 sensor cap is 1 year from the date of delivery. If the above products or parts within the warranty period is not functional due to raw material defects or poor manufacturing, they are free for repair or replace.
12 Trouble Shooting Error Solutions 1. The battery is not installed correctly. Check the direction. The instrument does 2. Battery low voltage, replace the battery. not turn on 3. Instrument freezes. Take out the batteries and re-install. 1. Check calibration procedure: correct atmospheric pressure, salinity input and temperature.
Salinity = Dissolved salts in water. Chlorinity = Measure of chloride content, by mass, of water. S (‰) = 1.80655 x Chlorinity (‰) 12 Appendix B: DO % Calibration Values Cal. Cal. Pressure Pressure value value D.O. in Hg mmHg mbar D.O.
Need help?
Do you have a question about the DO8500 and is the answer not in the manual?
Questions and answers