3
Service
9 Disconnect the DUT from the power meter 1 mW calibrator.
10 Terminate the DUT with the 50 MHz, 50 Ω load.
11 Perform zeroing for the DUT by sending "CAL:ZERO:AUTO ONCE".
12 Set the DUT to the single trigger mode by sending "INIT:CONT OFF".
13 Read the noise level of the DUT by sending "READ?" and then record
14 Repeat step 13 for 10 times and then calculate the mean value of the
15 Compare the calculated mean value to the system specification. If the
52
(This step is compulsory for N8488A but optional for N8481A, N8481B,
N8481H, N8482A, N8482B, N8482H, N8485A, N8487A, N8486AR, and
N8486AQ.)
the reading.
readings.
test fails, refer to
"Repair" on page
59.
N8480 Series Power Sensors Operating and Service Guide