Priming the Tubing and Starting
the Pump
When the new cassette is attached, the pump may automati-
cally prompt you to prime the tubing.
1. If Prime Tubing? appears,
´
press
.
2. This screen will appear. Make
sure you are not connected
to the pump's tubing. Open
the tubing clamps.
WARNING: Do not prime the fluid path with the tubing
connected to your catheter. This could result in over-delivery
of medication or air embolism, which could result in death or
serious injury.
3. Press and hold the
until the tubing is fully
primed or until priming
stops.
4. This screen will appear if you
release the
priming automatically stops:
If the tubing is not yet
fully primed, press the
´
key and repeat step 3.
If you are finished priming, press the
´
key
´
key or if
Prime Tubing?
Press Y or N
Disconnect tubing
from patient
Open clamps
Hold Y to prime
Priming...
0.1 ml
Hold Y to prime
Continue Priming?
Press Y or N
Î
key.
19