Removing A Cassette - Deltec CADD-Prizm VIP CONTIN 6100 Patient Information

Infusion pumps; continuous delivery mode
Table of Contents

Advertisement

R e m o v i n g a Cassette
WARNING: Close the tubing clamp before removing the
cassette from the pump to prevent unregulated infusion,
which could result in death or serious injury.
To remove a cassette:
1. Press
pump. When Stop the
pump? appears, press
2. Close all tubing clamps and discon-
nect the tubing from your access
device as instructed by your clini-
cian.
3. Use a coin to unlatch the cassette.
Insert the coin into the slot and
turn clockwise until the latching
button pops out.
NOTE: If you cannot turn the
latching button, the cassette
may be locked. Contact your
clinician.
4. Remove the cassette from
the pump.
5. Discard the Medication
Cassette
®
CADD
Administration Set
as instructed by your
clinician.
14
to stop the
´
Reservoir or
Stop the Pump?
.
Press Y or N
2000-03-08 D. Z
«Prizm Remove C
®
®

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cadd-prizm vip contin 6101

Table of Contents