Attaching A Cassette - Deltec CADD-Prizm VIP CONTIN 6100 Patient Information

Infusion pumps; continuous delivery mode
Table of Contents

Advertisement

A t t a c h i n g a Cassette
WARNING:
• You must use a CADD
Valve or a CADD
integral or add on Anti-Siphon Valve to protect against
unregulated gravity infusion that can result from an
improperly attached cassette. Unregulated gravity infu-
sion can result in death or serious injury.
• For detailed instructions and warnings pertaining to the
Medication Cassette
Set, please refer to the instructions for use supplied with
those products.
• Frozen medication must be thawed at room temperature
only. Do not heat the Medication Cassette
microwave oven as this may damage the medication, the
Medication Cassette
Use aseptic technique as
instructed by your clinician.
1. Clamp the tubing on the
new Medication
Cassette
Reservoir or
®
CADD
Administration
Set.
2. Insert the cassette hooks
into the hinge pins on
the pump.
®
Extension Set with Anti-Siphon
®
Administration Set with either an
Reservoir or CADD
Reservoir, or cause leakage.
®
Administration
Reservoir in a
2000-03-08 D. Z
«Prizm Hinge C
®
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cadd-prizm vip contin 6101

Table of Contents