Starting the Pump

When you start the pump, it will review the program, and
then begin delivering medication. If the pump will not start,
a message should appear on the display. Refer to the
Messages and Alarms section on page 23.
WARNING: Prior to starting infusion, inspect the fluid path
for kinks, a closed clamp, or other upstream obstructions,
and remove all air bubbles to prevent air embolism. An
undetected upstream occlusion may result in under- or non-
delivery of drug and, depending on the drug being adminis-
tered, could result in death or serious injury. Air embolism
can result in death or serious injury.
1. Press
2. When Start the pump? ap-
pears, press
Starting pump... will appear. The
pump will automatically review
and display the preprogrammed
settings.
RUNNING will appear on the
main screen, the green light will
blink, and fluid delivery will
begin as programmed by your
clinician.
12
.
´
.
Start the Pump?
Press Y or N
Starting pump...
***
CONTIN 6220X ***
Res Vol
Press NEXT to advance
RUNNING
50.0 ml