Trouble Shooting - Omegasonics OMG-1420BTD Operation & Instruction Manual

Table of Contents

Advertisement

Isolate the exact issues(s) you are experiencing by following the following procedure.
1. Turn off Ultrasound
2. Turn off Heat
3. Leave Green Power push button pressed in and illuminated.
4. Turn on Heat – Does it operate?
5. Turn off Heat
6. Turn on Ultrasound – Does it operate?
7. Turn off Ultrasound
8.
Control panel has lights, but there is no Heat action.
1. Is the water level to the middle height of the tank?
2. If NO - fill water to the appropriate level (14" – 16").
3. If this doesn't correct the problem:
a. Unplug the machine from the facility power supply.
b. Remove the panel located on the right end of the machine.
c. Look for the ice cube relay located at the left end of the back panel.
d. Flip the window up (window has a red button) – this bypasses the thermal
overload. You will see an orange flag visible in the upper right window.
e. The machine will be operational, but will have no low water level
protection.
4. If either instruction does not solve the issue, replace ice cube relay.
5. Contact Omegasonics to order a replacement ice cube relay.
No lights on the control panel.
1. Is the Power push button pushed in?
2. Is there power to the facility outlet?
3. Turn off Power and unplug the machine.
4. Remove upper panel located on the right end of the machine.
5. Is the circuit breaker energized (UP-Red) or (DOWN-Green)?
6. If the circuit breaker is tripped, re-set the circuit breaker.
7. Plug in machine.
8. Follow the steps above to isolate the exact issue.

TROUBLE SHOOTING

13