SECTION 5 – DATA MANAGEMENT
OVERVIEW
The AED is designed for ease of data management and review. The data stored in internal memory can be
displayed on the PC screen using the RescueLink software.
Topic

Recording Rescue Data

Review Rescue Data
RECORDING RESCUE DATA
RECORDING DATA IN INTERNAL MEMORY
The AED automatically stores up to 34 minutes of rescue data.
If the internal memory contains a previously stored rescue when a rescue is attempted, the voice prompt
"Data in Memory. Push Flashing Button to Erase Data and Perform Rescue" will be heard. Pressing the
CONTINUE button will erase the data and allow the rescue attempt to continue.
Note: Do not press the CONTINUE button unless you are sure you want to erase the internal
memory in the AED.

REVIEWING RESCUE DATA

RETRIEVING DATA FROM MEMORY
1.
Open the AED lid.
2.
Connect the serial cable to the PC and to the AED's serial port under the blue rubber data access
cover. The voice prompt will say " Communications Mode."
3.
Run the RescueLink software program.
4.
Select COMMUNICATIONS, GET RESCUE DATA. On the RescueLink software program.
6.
Select INTERNAL MEMORY OF AED then select OK.
WARNING: Electric Shock and Fire Hazard
Do not connect any telephones or unauthorized connectors to the socket on this equipment.
CAUTION: Serial Communication Cable
The serial communication cable is only for use with the AED; it is not to be used with a telephone.