With access to the Andalink, you can easily manage all your Soberlive breathalysers from multiple sites as well as get test reports sent to your email, receive notifications and change user information and login details online.
With Andalink, company and OH&S managers can register users, as well as view, download, and manage reports for their employees’ alcohol results. For instructions on how to set up your Andalink account for your Soberlive, please visit: andatech.com.au/andalink-manual...
Soberlive and plug into a power outlet. 4. Use provided screws to attach the Soberlive to the wall. 5. Close the Soberlive unit and lock the cover into place. Ensure to keep the key in a secure place.
Page 5
Date, Time and Wi-Fi Important Note to Administrators / Managers: Ensure that the time and date on the Soberlive is always set correctly. 1. Tap on the Wi-Fi icon. 2. The Wi-Fi screen will show the current date and time of the device on the top right of the screen.
Shows all the test results taken on the device. Manage User ADD NEW USER - Add a new user for this Soberlive device. SYNC - Tap this button to sync all the users between the Andalink and the device. (Note: Make sure that the device is connected to the internet.) GUEST LOGIN - Switch the guest login on or off.
Used when replacing sensors for calibration. Refer to Replacing Sensors section. Email Settings Customise the Soberlive email settings for the test reports. Sync with server - This option will sync the most recent settings with server. (Make sure the Soberlive is connected to the internet) Allow Emailing - This option will enable or disable email features.
Measurement Unit Choose the preferred measurement unit. Tap on your selection and tap on "Back" to save. High Low Settings Choose the High and Low level for this device. If the user exceeds the limit, it will trigger a continuous beep sound after taking the test.
Page 9
Set the time to turn the device on and off automatically. Precaution: If device does not wake up on the selected time. just turn off the power supply to the Soberlive and turn it back on after 5 seconds. Update Andatech may update the device software from time to time.
Page 10
3. Tap on "Install" 4. The update will start installing. 5. Once installed, tap on "Open" to complete the update. Admin Exit Tap this to log out of admin /manager view. The Soberlive will return to the login screen.
3. Remove the top and bottom screws attaching the sensor to the Soberlive. Sensor 4. Pull the sensor out of the Soberlive. 5. Replace with a newly calibrated sensor. Write down the new sensor's serial number. 6. Replace and tighten the top and Screw bottom screws.
Page 12
11. Another notice for calibration may appear after login. Click "OK" to close the notice and load the admin panel. 12. Click the Sensor Swap icon from the admin panel. 13. Enter the new sensor's serial number and then tap "Apply Setting" to save.
Page 13
OPERATING GUIDE 1. Log in to a user account on the Soberlive using an ID and PIN. For guest users, just click "GUEST LOGIN" to take a test (no ID and PIN is required). Note: This option will only appear if "Allow Guest" is enabled in the Location or Soberlive settings.
There is a hardware problem on the PCB. Contact customer service. 5. This status is not T2. Please check Analyzer status T2 indicates that the tablet PC is ready to receive data from the Soberlive. If you see this error message, contact customer service.
Page 15
Soberlive's admin panel. Once logged in, managers cannot take a test. A manager must have set up their own user ID and PIN in order to take a test like a regular user. For users, their PIN and ID can be used on the Soberlive only, not for the Andalink.
Page 16
DISCLAIMERS • This product must be used strictly in accordance with the instructions in this manual. • This product is intended to provide the user with an indication of alcohol on the breath. However, neither the manufacturer nor the supplier gives any warranty (express or implied) that the product can give any indication of blood alcohol level.
Page 17
The liability for proper functioning of the instrument is irrevocably transferred to the owner or operator in the event the instrument is serviced or repaired by personnel not employed or authorised by Andatech. In such an event, the warranty will be void. WARRANTY CALIBRATION AND SERVICE...
To claim for warranty, contact us on 1300 800 200 or send and email to support@andatech.com.au to get an RA# (Return Authorisation Number). Ensure that the RA# is clearly stated on the outside of the packaging and that the product is properly packaged so that no damage occurs to the product during transit.