Andatech AlcoSense PRODIGY II User Manual

Andatech AlcoSense PRODIGY II User Manual

Industrial breathalyser

Advertisement

Andatech
®
P R O D I G Y II
Industrial Breathalyser
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AlcoSense PRODIGY II and is the answer not in the manual?

Questions and answers

Summary of Contents for Andatech AlcoSense PRODIGY II

  • Page 1 Andatech ® P R O D I G Y II Industrial Breathalyser USER MANUAL...
  • Page 2: Table Of Contents

    INTRODUCTION Thank you for purchasing the Andatech Prodigy II. ® The Andatech Prodigy II is a state-of-the-art instrument used to detect ethyl alcohol from a valid breath sample. ® It uses an electrochemical fuel cell to measure the concentration of alcohol in the breath sample expelled from a human subject.
  • Page 3: Device Features

    DEVICE FEATURES DIAGRAM EXPLANATION Battery Charging Stylus Indicator/Mouthpiece Sample Port (back of unit) Inserting Indicator Button to Remove Touch Screen Mouthpiece On/Off Function Button Button Printer 1. DISPOSABLE MOUTHPIECE 4. TOUCH SCREEN DISPLAY Each mouthpiece is individually packed and sealed for hygiene This displays the instructions and the alcohol concentration reasons.
  • Page 4: Conditions For Testing

    Andatech. In such an event, the warranty will be void. There must be an interval of at least 15 minutes after •...
  • Page 5 If no data fields are selected, the unit will proceed directly to test mode. (Contact Andatech to enable/disable data fields) 6. When the screen displays “Please Blow”, take a deep breath and blow into the mouthpiece with a slow and steady breath until the BEEP tone stops.
  • Page 6: Test Records

    In order to download your data to a PC you must have the Andalink for Prodigy II software, available from 5. Select the test record number to view the test www.andatech.com.au record. SETTINGS 1. Auto Off iii. Uncheck the “Print GPS” tickbox to disable printing the Longitude and Latitude on test i.
  • Page 7: Touch Screen Calibration

    If the touchscreen is not sensitive enough or not responding, calibrate the screen. Follow the instructions on page 11. AIR BLANK TEST CONTINUES TO DETECT ALCOHOL Let the device sit for 5 minutes to clear out any possible residue. If the problem persists, contact Andatech.
  • Page 8: Disclaimers

    DISCLAIMERS This product must be used strictly in accordance with the instructions in this manual. This product is intended to provide the user with an indication of alcohol on the breath. However, neither the manufacturer nor the supplier gives any warranty (express or implied) that the product can give any indication of blood alcohol level.
  • Page 9: Warranty Details

    Service during the Warranty Period To claim for warranty, contact us on 1300 800 200 or support@andatech.com.au to get an RA# (Return Authorisation Number). Ensure that the RA# is clearly stated on the outside of the packaging and that the product is properly packaged so that no damage occurs to the product during transit.
  • Page 11: Product Registration

    PRODUCT REGISTRATION Visit my.andatech.com.au to register your warranty online and receive calibration reminders. Alternatively, complete this form and forward to the address below with a copy of your proof of purchase: Andatech PO BOX 3038, Nunawading VIC 3131 Australia. PRODUCT: ANDATECH PRODIGY II ®...
  • Page 12 Andatech Pty Ltd www.andatech.com.au 1300 800 200 Version 5. Updated 10.11.2017.