AgaMatrix JAZZ DoseCoach Owner's Manual page 223

Dose helper and blood glucose meter
Hide thumbs Also See for JAZZ DoseCoach:
Table of Contents

Advertisement

Summary of total number of acceptable results per EN ISO 15197:2015
Total (percent)
581/600 (96.8%)
All results were within the clinically acceptable A and B zones of the Parkes Error
Grid with 100% falling within zone A.
Acceptance criterion of 95% within ± 0.83
mmol/L and within ± 15%
Meets Criterion
221

Advertisement

Table of Contents
loading

Table of Contents