Performing a control solution test
Do
4
4
Control solution should only
be used for 90 days after first
opening the bottle or until the
date of expiry printed on the
label, whichever comes first.
4
4
You should only use AgaMatrix
control solution with JAZZ™
DoseCoach
4
4
Always put the cap on the
control solution bottle right
after you have used it.
4
4
Do not add water or any other
liquid to the control solution.
4
4
Control solution tests should
be performed within the
temperature range of 10°C to
40°C.
154
.
®
The control solution is used to:
Make sure that the AgaMatrix
WaveSense JAZZ test strips work with
JAZZ™ DoseCoach
Practice measuring without having to
use your own blood.
Control solution tests should be
done when you:
First get your JAZZ™ DoseCoach
Think your JAZZ™ DoseCoach
or test strips are not working
properly.
Think your readings are
inaccurate.
Have dropped or damaged your
JAZZ™ DoseCoach
such as water on it.
Are told by your healthcare
professional to do so.
.
®
.
®
®
or got liquids
®