MECHANICAL INSTALLATION

It is recommended to mount Cava™ at approximately 5½' to 6' from the floor. This will allow the interface to be at an
appropriate height for use during setup and during normal operation.

Mounting Options

Cava™ can be mounted in one of three ways. Option A uses a pre-cut single gang junction box cut-out. Option B uses an on-
site created cut-out using a Jigsaw. Option C uses an on-site created circular cut-out.
CAUTION
There are often disconnect switches and gas/water lines running in the fume hood. Make sure that the marked location for
the interface does not conflict with any of the other components that may be installed on/in the fume hood. When cutting
and drilling holes in the fume hood, ensure proper precautions are taken to avoid scuffing or scratching the fume hood.
3 | CAVA – Manual |
AntecControls.com
Option A
Mount Cava™ using a pre-cut single gang junction box cut-out.
STEP 1
Mount the backplate to the fume hood using four #6 self-tapping metal screws (minimum 3/8"
long).
STEP 2
Attach Cava™ and tighten the set screw once electrical installation is complete.
Option B
Mount Cava™ using a Jigsaw cut-out.
STEP 1
Mark the appropriate mounting location for Cava™.
STEP 2
Drill two 1/4" to 1/2" pilot holes at either corner of the provided template in
STEP 3
Using a jigsaw, cut four lines to cut-out a square.
STEP 4
Drill four 7/64" mounting holes according to the provided template in
STEP 5
Deburr and file any sharp edges if they are present to prevent any edges from cutting the
wiring.
STEP 6
Mount the backplate to the fume hood using four #6 self-tapping metal screws (minimum 3/8"
long).
STEP 7
Attach Cava™ and tighten the set screw once electrical installation is complete.
Option C
Mount Cava™ using a circular cut-out. This is the least preferable mounting option, as the
hole provides the least access to the wiring terminals.
STEP 1
Mark the appropriate mounting location for Cava™.
STEP 2
Drill a 2.25" hole in the front panel of the fume hood according to the provided template in
Appendix
A.
STEP 3
Drill four 7/64" mounting holes according to the provided template in
STEP 4
Deburr and file any sharp edges if they are present to prevent any edges from cutting the
wiring.
STEP 5
Mount the backplate to the fume hood using four #6 self-tapping metal screws (minimum 3/8"
long).
STEP 6
Attach Cava™ and tighten the set screw once electrical installation is complete.