Configurable Fume Hood Modes

Cava™ comes with four fully configurable fume hood modes. These modes can be configured and updated through Antec
Toolbox by using the RIN Ethernet Port on the back of the device.
The fume hood modes will be configured during system start-up by your local certified start-up technician. Refer to your
NOTE:
start-up technician for changes to your fume hood modes.
Antec Toolbox Fume Hood Mode Options
Mode Option
Available Options
Name (Global)
19 Characters
Control Method
Face Velocity
Constant Volume
Full Open
Full Close
Priority (Global)
First
Second
Third
Default
Mode Entry Delay
No Limits
Mode Exit Delay
No Limits
Color (Global)
Green
Red
Blue
Yellow
Trigger Alarm
On
(Global)
Off
Global mode options are common between each Cava™ in a room to provide consistency between fume hoods.
NOTE:
Alarms
20 | CAVA – Manual |
AntecControls.com
Description
Set the name of the mode that will be displayed on CAVA and reported to BACnet
Determines the method with which CAVA will control the fume hood.
NOTE:
Face Velocity requires a face velocity setpoint, minimum flow, and maximum flow.
NOTE:
Constant Volume requires a target flow.
Determines which mode will be entered when multiple modes meet their entry conditions. Default
has the lowest priority.
Number of seconds CAVA will wait to enter mode while entry condition is met.
NOTE:
This is not configurable in the Default priority mode.
Number of seconds CAVA will wait to exit mode while exit condition is met.
NOTE:
This is not configurable in the Default priority mode.
Background color on CAVA Home Screen
Set whether an alarm will be triggered when the mode is entered.
Alarm Notification Screen
If any alarms are triggered, an alarm notification screen will appear and indicate the cause of
the alarm. The user can also trigger an alarm at any time using the large tactile emergency
button.
Silencing the Alarm
When an alarm notification screen appears, the user will have the option to silence the alarm
for a configurable amount of time (see
gloves, the large tactile mute button can also be used to silence the alarm.