CAUTION
Mounting position is critical to ensure the accuracy of the SVS.
Ensure that there are no sharp bends or kinks in the tubing during
the installation. Improper installation will cause failure of the SVS.
Regardless of where the SVS is mounted, the following steps outline how to install the SVS components.
16 | CAVA – Manual |
AntecControls.com
STEP 1
With the sash in the fully open position, drill a 5/16" hole 5" above and 5" inward from the
base of the sash.
STEP 2
Drill a 5/16" hole on the front of the fume hood on the same side as the first hole.
STEP 3
Mount the brass bulkheads to the fume hood through the holes.
STEP 4
Select one of the three mounting options on the previous page for the SVS and mount using
one of the two methods outlined below:
For 3M Dual Lock Fastener mounting (recommended), clean the mounting surface and
the back of the sensor with a cloth or IPA. Wait 5 minutes for IPA to evaporate if used. At
room temperature, peel backing from one side of the 3M fastener and apply it to the back of
the sensor. Peel backing from opposite side of the 3M fastener and firmly press the SVS and
3M fastener onto the mounting surface for 30 seconds.
For screw mounting, place sensor in desired location and mark the three mounting hole
locations on the fume hood. Drill three 1/8" pilot holes in the appropriate locations on the
fume hood. Mount sensor using three #8 x 5/16 in. screws.
STEP 5
Use the high-pressure tubing to connect the outer brass bulkhead to the high-pressure port
of the sensor. Do the same to connect the inner brass bulkhead to the low-pressure port of
the sensor.
CAVA - MANUAL