INTERFACE

Home Screen

The Cava™ Home Screen will be displayed on the interface once setup has been completed through Antec Toolbox. The Home
Screen provides the user with a clear indication of the fume hood status and face velocity reading.
Four screens containing additional information regarding the functionality and state of the fume hood can be accessed from the
home screen by swiping left. The user can return to the home screen at any time using the home button in the top right corner.
1
Displays the control method, fume hood mode, face velocity target, and current sash open percentage if a sash position sensor is connected.
2
Displays the current face velocity, current valve airflow, current valve pressure (when used with Venturi Valves), and current actuator voltage.
3
Displays the sash height of any sash position sensors wired into CAVA.
4
Displays information about the Cava™ device and can be accessed through the info button in the bottom right corner.
19 | CAVA – Manual |
AntecControls.com
A
B
C
D
E
F
1
Home Screen Components
Display Component
A
Settings Menu
B
Fume Hood Name
C
Fume Hood
Status
D
Face Velocity
E
Mute Button
F
Emergency
Button
2
Description
Allows access to settings under password
protection. Default password is 1-6-6-4 and can be
changed using Antec Toolbox.
The current name of the fume hood is displayed.
This can be changed using Antec Toolbox.
Indicates the status of the fume hood and any
special conditions that may be present.
Displays current fume hood face velocity.
Used to temporarily silence an alarm for a
configurable number of minutes (see
Preferences
section).
Used to quickly enter emergency mode, triggering
an alarm. Once triggered, it can be disabled by
pressing the Emergency Button for 5 seconds.
3