WatchGas UNI SUSTAINABLE Quick Start Manual

WatchGas UNI SUSTAINABLE Quick Start Manual

Single gas detector
Hide thumbs Also See for UNI SUSTAINABLE:

Advertisement

Quick Links

D e t e c t i o n
Quick Start Guide
UNI Sustainable Single Gas Detector

Advertisement

Table of Contents
loading

Summary of Contents for WatchGas UNI SUSTAINABLE

  • Page 1 D e t e c t i o n Quick Start Guide UNI Sustainable Single Gas Detector...
  • Page 2: Read Before Operating

    Remove the monitor cover and battery only in area known as non-hazardous. • Use only WatchGas lithium battery part number M500-0001- 000 [1.17.02.0002] (3.6 V, 2700 mAh, AA size). • This instrument has not been tested in an explosive gas/air atmosphere having oxygen concentration greater than 21%.
  • Page 3 User Interface The UNI’s user interface consists of the LCD display, LEDs, an alarm buzzer, two keys: Left Key [Confirm/ Number increasing] and Right Key [Power/Cursor moving], an alligator clip, and a chemical sensor. ALLIGATOR CLIP LED ALARM WINDOW LCD DISPLAY BATTERY STATUS P O W E R / C U R S O R CONFIRM/...
  • Page 4 Display Detail Gas name, including CO, H2S, SO2, or O2 Question mark (perform operation?) OK status indicator (or accept number entry) Concentration unit, e.g. ppm, %, mg/m3, etc. Battery charge status Alarm type indicator Zero in progress indicator Calibration in progress indicator Concentration or other numerical value...
  • Page 5: Normal Mode

    Turning the Unit On Press and hold the Right Key for 4 seconds, until the LCD displays “On”, the buzzer beeps, and the green LED is on. The unit enters a warm up and self-test sequence, shows the firmware version, e.g. “VER”, ”...
  • Page 6: Configuration Mode Menu

    Configuration Mode Menu In Config Mode the user can do calibrations and chan- ge various operating parameters: • AIR, SPAN: Zero or Span calibration, • BUMP: Bump test • SET HIGH\LOW\STEL\TWA alarm limits • SET CAL: Change span value • CAL INTV, BUMP INTV span or bump interval •...
  • Page 7: Alarm Limits

    Alarm Limits To adjust an alarm limit, enter Config Mode and scroll to the desired alarm: SET HIGH\LOW\STEL\TWA?. • Press Left to display the alarm value with the first digit flashing • Use Left to move to the desired flashing digit •...
  • Page 8 Span calibration uses a known concentration gas to determine the response of the sensor to the gas. Attach the Calibration Adapter over the inlet port on the front of unit by pressing it into place. Use a fixed-flow regulator with a flowrate of preferably about 0.3 to 0.5 LPM.
  • Page 9: Maintenance And Service

    Maintenance and Service If used in high-dust environments, replace the inlet filter regularly. Battery and sensor can be replaced as needed. For other maintenance, contact mPower’s service department or refer to the full User Manual. REMOVE 4 TURN AROUND REMOVE UNIT COVER SCREWS DETECTOR REPLACE RUBBER...
  • Page 10 Please use the return and collection system available in your country for the disposal of this product. D e t e c t i o n WatchGas Sextantstraat 61 2901 ZZ Capelle aan den IJssel +31 (0)85 01 87 709 V.1.0...