Corrective Maintenance
5.2
dISaSSeMBly/reaSSeMBly (Continued)
5.2.6 Internal Cable (handheld Connector)
and lens
1.
Use a socket wrench to remove Internal
Cable nut from inside Front Case
Assembly.
2.
Remove washer and then pull Internal
Cable from Front Case.
3.
Remove screws (
Front Case and remove Lens.
4.
Remove Gasket from Front Case
opening.
Nut
5-10
Alaris
Auto-ID Module
®
Technical Service Manual
) attaching Lens to
2
Gasket
Lens
during reassembly:
1.
When installing Gasket, ensure it is
positioned within window frame.
2.
When installing Lens:
a. Wear finger protection.
b. If it is a new Lens, remove protective
c. Prior to tightening screws, gently tap
CauTIOn
Use care when tightening the screws
(see Table
lens.
d. Ensure Lens is clean.
Washer
film from Lens.
lens to ensure it is seated on Gasket.
) to avoid cracking the
5-2