Iui Connectors And Case Separation - CareFusion Alaris 8600 Technical & Service Manual

Auto-id module
Hide thumbs Also See for Alaris 8600:
Table of Contents

Advertisement

Corrective Maintenance
5.2
dISaSSeMBly/reaSSeMBly (Continued)

5.2.2 IuI Connectors and Case Separation

1.
Remove screws (
(left and right) to module.
2.
Remove
IUI
3.
Remove screws (
washers attaching Rear Case to Front
Case Assembly.
4.
Pull Rear Case away from Front Case
Assembly.
5.
If replacing Rear Case, note model/
reference number and serial number.
5-4
Alaris
Auto-ID Module
®
Technical Service Manual
) attaching each
2
Connectors and Gaskets.
) and associated
2
during reassembly:
• Ensure ground clips are still installed on
IUI
both
• To install Right
remove protective backing and adhere to
IUI
• To install Left
seal on one end of connector and stretch
to other end to conform to connector
body. Gently press on seal to seat
completely. Use lint-free swab to apply
alcohol to top, sides, and bottom of seal
for lubrication while installing to Rear
Case. Do not apply alcohol to contacts or
connector.
• Ensure Silicone Tubing in Rear Case is in
place and not damaged.
• Use care not to pinch wires when
reinstalling Rear Case.
• If replacing Rear Case:
a. Use permanent black ink to print
b. Apply Serial Number Replacement
c. Adhere transparent label over Serial
Connectors.
IUI
Connector Gasket,
IUI
Connector.
Connector Seal, position
IUI
instrument's model/reference number
and serial number on Serial Number
Replacement Label.
Label to instrument.
Number Replacement Label.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents